ZHU Tuo (朱托), SHANG Wanli (尚万里), ZHANG Wenhai (张文海), YANG Jiamin (杨家敏), et al.. Quantitative Measurement of the Proportions of High-Order Harmonics for the 4B7B Soft-X-Ray Source at Beijing Synchrotron Radiation Facility[J]. Plasma Science and Technology, 2013, 15(12): 1194-1196. DOI: 10.1088/1009-0630/15/12/06
Citation: ZHU Tuo (朱托), SHANG Wanli (尚万里), ZHANG Wenhai (张文海), YANG Jiamin (杨家敏), et al.. Quantitative Measurement of the Proportions of High-Order Harmonics for the 4B7B Soft-X-Ray Source at Beijing Synchrotron Radiation Facility[J]. Plasma Science and Technology, 2013, 15(12): 1194-1196. DOI: 10.1088/1009-0630/15/12/06

Quantitative Measurement of the Proportions of High-Order Harmonics for the 4B7B Soft-X-Ray Source at Beijing Synchrotron Radiation Facility

  • A transmission grating coupled with an X-ray charge coupled device (CCD) is used to quantitatively measure the proportion of high-order harmonics of the soft-X-ray source of beam line 4B7B. The results show that the monochromatic X-ray has third-order and second-order harmonics. The proportion of second-order harmonic of 4B7B is less than 9.0% and the third- order harmonic is below 0.7% when no suppressing method is applied. When suppression methods are used, the proportion of second-order harmonic is less than 1.7% and the third-order harmonic is ignorable.