Yuan YAO, Yao YANG, Ang TI, Yang SONG, Jiamin ZHANG, Yan WANG, Yao ZHANG, Haiqing LIU, Yinxian JIE. A fringe jump counting method for the phase measurement in the HCN laser interferometer on EAST and its FPGA-based implementation[J]. Plasma Science and Technology, 2024, 26(4): 045601. DOI: 10.1088/2058-6272/ad0dec
Citation: Yuan YAO, Yao YANG, Ang TI, Yang SONG, Jiamin ZHANG, Yan WANG, Yao ZHANG, Haiqing LIU, Yinxian JIE. A fringe jump counting method for the phase measurement in the HCN laser interferometer on EAST and its FPGA-based implementation[J]. Plasma Science and Technology, 2024, 26(4): 045601. DOI: 10.1088/2058-6272/ad0dec

A fringe jump counting method for the phase measurement in the HCN laser interferometer on EAST and its FPGA-based implementation

More Information
  • Author Bio:

    Yao YANG: yangyao@ipp.ac.cn

  • Corresponding author:

    Yao YANG, yangyao@ipp.ac.cn

  • Received Date: August 04, 2023
  • Revised Date: November 16, 2023
  • Accepted Date: November 16, 2023
  • Available Online: April 08, 2024
  • Published Date: April 10, 2024
  • Electron density in fusion plasma is usually diagnosed using laser-aided interferometers. The phase difference signal obtained after phase demodulation is wrapped, which is also called a fringe jump. A method has been developed to unwrap the phase difference signal in real time using FPGA, specifically designed to handle fringe jumps in the hydrogen cyanide (HCN) laser interferometer on the EAST superconducting tokamak. This method is designed for a phase demodulator using the fast Fourier transform (FFT) method at the front end. The method is better adapted for hardware implementation compared to complex mathematical analysis algorithms, such as field programmable gate array (FPGA). It has been applied to process the phase measurement results of the HCN laser interferometer on EAST in real time. Electron density results show good confidence in the fringe jump unwrapping method. Further possible application in other laser interferometers, such as the POlarimeter-INTerferometer (POINT) system on EAST tokamak is also discussed.

  • Electron density is one of the most important parameters in the magnetically confined fusion plasmas. Measurement of the refractive index (N) of plasma by using the electromagnetic wave is a well-established tool for measuring the electron density for many decades [16]. Far-infrared (FIR) laser interferometer detects the line-integrated electron density through launching a beam of FIR laser into the plasma. Because the refractive index of plasma is different from that in the vacuum condition (N = 1), some phase difference (corresponding to optical path difference) will generate between the probing beam and the reference beam without passing through the plasma. Based on the interferometry technique, we can achieve the line-integrated electron density by: φ(rad)=kλnedl, here, λ is the wavelength of electromagnetic wave, ne is the local electron density, dl is the line element of integral operation, the coefficient k=(e2/4πc2ε0me)=2.82×1015, where e is the elementary charge, c is the light speed, ε0 is the dielectric constant of vacuum, me is the mass of electron.

    HL-3 is a newly constructed tokamak with major radius of 1.78 m and minor radius of 0.65 m [7, 8]. HL-3 will focus on high parameter plasma operation and physics research, such as high confinement mode (H-mode), energetic particles driven instabilities, advanced divertor physics, and so on. Aiming to measure the electron density and perform real-time electron density feedback control, a multi-channel formic-acid laser (HCOOH, λ = 432.5 μm) Michelson-type interferometer was proposed on HL-3, including eight horizontal channels and five oblique channels. Five horizontal channels distributed in the central region (corresponding to the geometry location Z = −20, −10, 0, 10, 20 cm) had been successfully developed in the early 2023. Subsequently, in order to investigate the boundary electron density activity during the divertor discharge with elongation, other three horizontal interferometry channels located at Z = −97, −76, 76.5 cm were recently developed on HL-3.

    This paper focuses on the development of the boundary three-channel FIR laser interferometer on HL-3. Section 2 describes the optical design and layout. Section 3 introduces the main components of the interferometry system. Section 4 represents the recent experimental result on HL-3.

    The boundary interferometry system shares the same laser source, front light path, support tower and inner space with the existing five-channel HCOOH laser interferometer, and achieves the electron density measurement through beam splitter. Optical design of the three boundary interferometry channels is based on the Gaussian beam propagation and lens transformation techniques [912].

    In principle, as a Gaussian beam with wavelength of λ propagates through the space, its light spot will expand and the diameter (d) can be determined by:

    d=[d20+(4λ2/π2)(L2/d20)]1/2 , (1)

    here, L indicates the distance from the beam waist (d0).

    Due to long distance propagation in the interferometry system, it is necessary to employ concave mirror with proper focal length (f) for beam transformation, as shown in figure 1. The beam waist parameter (d2, L2) after beam transformation can be determined from original beam waist (d1, L1) as follows:

    Figure  1.  Principle of the lens transformation for Gaussian beam.
    d2={1/[(1L1/f)2/d12+(πd1/2λ)2/f2]}1/2, (2)
    L2=(L1f){f2/[(L1f)2+(πd12/(2λ))2]}+f. (3)

    Figure 2 shows the optical layout of the HCOOH laser interferometer on HL-3. For the convenience of description, the light paths of central five horizontal channels are also provided in figure 2 by the light color. As the probing wave and local oscillation (LO) wave come out of the long waveguide pipes (~ 19 m), they get into the main optical path. In order to isolate possible vibration from the ground and HL-3 device, all the optical components are assembled on the big optic board which is firmly installed on the large stainless-steel support tower (7.0 m in height, ~ 8 tons in weight) next to the No. 5# diagnostic window of HL-3 tokamak.

    Figure  2.  Optical layout of the three boundary interferometry channels (red color), together with the central five horizontal channels (light color).

    At the bottom of support tower, two plane mirrors (M1-1, M2-1) are accurately assembled to reflect the probing wave and LO wave into the main optical layout. In order to derive the reference beat signal (R), firstly, small parts of probing beam and LO beam are individually separated out by the mesh-grid splitters, and then combined at the second mesh-grid. Finally, the combined waves are received by the high responsivity Schottky diode detector (DR), generating the beat signal (R).

    For the probing beam of the three boundary interferometry channels, the design of light path is based on existing central five interferometry channels through adding mesh-grid splitters with proper transmission/reflection ratio. For the two lower boundary channels, their probing waves are obtained through separating some light-intensity from previous central channel 1# and channel 3#, respectively. For the upper boundary channel, its probing wave is obtained through separating some light-intensity from central channel 5#. Similarly, the LO waves of the three boundary interferometry channels are achieved through separating wave from central channels 1#, 3#, 5#, respectively. The probing waves are orderly launched into HL-3 plasma along Z = −97, −76, 76.5 cm, and reflected by the metal retro-reflector. Because the boundary interferometry channels are close to the plasma edge, where the electron density is very low, the refraction effect caused by the density gradient is weak. Meanwhile, the size of reflectors is big enough to receive the probing beams from plasma. Finally, the probing wave carrying the information of electron density combined with the outside LO waves, and received by the Schottky diode detectors (D1, D2, D3), generating the probing beat signals (P1, P2, P3). Finally, the line-integrated electron density can be achieved by computing the phase difference between P and R.

    The FIR laser interferometer on the HL-3 tokamak shares the terahertz formic acid (HCOOH) laser source with HL-2A tokamak [13, 14]. In practical operation, two FIR laser diagnostics can be freely switched by a removable plane mirror which is mounted close to the lasers. HCOOH laser is optically pumped by a high-power CO2 laser (~ 50 W). A Fabry-Perot (FP) cavity is employed to stabilize the CO2 pump laser’s frequency for optimal FIR pumping. The process involves sending the laser through the FP, detecting the transmitted signal with a phase-sensitive detector (PSD), and then the length of the FP is dithered about the nominal point. The PSD output reflects the frequency difference between the FP and CO2 pump laser.

    The HCOOH laser emits linearly polarized Gaussian radiation with a wavelength of 432.5 μm, beam waist of 11.0 mm and output power level of 15–25 mW in operation. Two HCOOH lasers are utilized here for heterodyne detection, and beat intermediate frequency (IF) in several megahertz can be obtained by slightly tuning the cavity length of HCOOH laser, so that the time resolution of (< 1.0 μs) can be achieved for the interferometry system.

    The newly developed three boundary interferometry channels are mainly used to analyze the electron density activity close to plasma boundary region during the divertor discharge with large elongation. Limited by the spatial structure of HL-3 tokamak around No. 5# section, only three probing channels are considered, in which one channel is located at the upper windows (ϕ300 mm) and two channel are located at the lower window (ϕ300 mm), respectively.

    Figure 3 shows the lower diagnostic window as an example. To ensure the minimum loss of light intensity at the diagnostic window, the aperture size (ϕ65 mm) is chosen to satisfy the design requirement. Aiming to withstand the high temperature baking on HL-3 (up to 300 °C), the FIR diagnostic windows in the boundary region are specially designed with an extended neck of 150 mm long. As a result, the temperature can fast drop at the position of crystal quartz plate and fluor rubber ring can be used for vacuum seal. In addition, to avoid the disturbance of reflected wave from the crystal quartz plate, three diagnostic windows are specially designed with 3.0° slant angle. Figure 3 gives the designed structure and picture of two lower diagnostic windows (ϕ65 mm).

    Figure  3.  Design and picture of the lower diagnostic window (ϕ300 mm) for the HCOOH laser interferometer on HL-3.

    In the Michelson-type interferometry configuration, the probing wave doubly passes through HL-3 plasma, so that the in-vacuum retro-reflector is required for beam reflection. On HL-3, three corner-cube-reflectors (CCR) which are made of copper with gold-plating are used for the boundary interferometry channels. CCR consists of three mutually perpendicular plane mirrors [15, 16]. Due to optic characteristic of the CCR, the reflected wave can propagate along the same path of incident wave by controlling the beam waist at the central area of CCR. These CCRs have an effective aperture of 50 mm, and all dihedral angle tolerances are less than 20". Figure 4shows the picture of CCR on HL-3 and primary test by using the visible laser (λ = 635nm). In the desktop test, visible laser orderly passes through the splitter and diaphragm, and then launches to the center of CCR which is ~ 3.0 m away from the diaphragm. The experimental result indicates that the reflected laser beam centrally crosses the diaphragm and can be observed below the splitter, which proves that the incident laser is perfectly reflected by the CCR along the same path.

    Figure  4.  Desktop test of the CCR by using the visible laser.

    The data processing system completes the phase difference computation and fast data acquisition. Figure 5 briefly displays the workflow of the data processing system, which starts to work as soon as receiving the trigger signal from HL-3 master control system. Firstly, the probing beat signal (P) and reference beat signal (R) output from detectors pass through band-pass filter and then are treated by 16-bit analog-to-digital converter (ADC) with sampling frequency of 10 MHz. Here, Field Programmable Gate Array (FPGA) is the key module used to compute the phase difference between P and R, based on the digital phase comparator technique [1719]. The raw beat signals and phase data are temporarily saved in the DDR-III and then uploaded to HL-3 database through fast PXIe bus for offline data analysis. Besides, the computed phase data is transmitted to a digital-to-analog converter (DAC) and delivered to HL-3 master control system for real-time density feedback control with a time delay < 1.0 ms.

    Figure  5.  Workflow of data processing system of the FIR laser interferometer.

    In the 2023 experimental campaign on the HL-3 tokamak, the newly developed three boundary interferometry channels were firstly put into operation and successfully measured effective electron density data. During the divertor discharge, the boundary interferometry channels are effective for investigating the electron density activity near the edge of plasma.

    Figures 6(a)–(c) give an example of experimental measurement on HL-3. In this shot No. 2164#, the plasma current is ~ 300 kA and the line-integrated electron density is ~ 1.5 × 1019 m−2 during the plasma flat-top phase. From the line-integrated electron densities shown in the figures 6(a)–(d), we can see that both boundary and central interferometry channels are sensitive to the supplementary gas puffing. In comparison, the evolution processes of electron density in figures 6(a)–(c) show more details due to the change of plasma configuration. The value of line-integrated electron density is generally smaller than 0.3 × 1019 m−2, which is about one order smaller than that in the plasma central region. For the upper channel shown in figure 6(a), the electron density shows bigger and faster change before 500 ms and then decreases. In the following phase, it gradually increases induced by gas puffing. On the contrary, the electron density measured by the lower two channels shows smaller before 400 ms and then increases in figures 6(b) and (c). In particular, figure 6(c) clearly reflects the density modulation process, and keeps steady level during the time of 400–1500 ms. For the three boundary channels, the channel (Z = −76 cm) shows higher electron density than other two channels (Z = −97, 76.5 cm), due to themselves detected positions and plasma vertical displacement which is shown in figure 6(e) during the divertor discharge. Figure 6(f) represents the plasma current (Ip), Dα emission (DVS- Dα) of the divertor region, and the gas puffing whose position is at the mid-plane.

    Figure  6.  (a)–(c) Temporal evolution of the line-integrated electron density measured by the boundary interferometry channels (Z = 76.5, −97, −76 cm), (d) line-integrated electron density measured by the central channel (Z = 0 cm), (e) plasma vertical displacement, (f) plasma current, gas puffing signal and Dα emission in the divertor region.

    Measurement of the electron density is very crucial for the plasma discharge operation and understanding of plasma physics. In order to investigate the boundary electron density activity during the divertor discharge on HL-3, three boundary interferometry channels located at Z = −97, −76, 76.5 cm have been firstly developed in 2023, based on the existing central five-channel HCOOH laser interferometer.

    In recent HL-3 experimental campaign, the line-integrated electron density of three channels is able to steadily operate with a time resolution of < 1.0 μs, line-integrated electron density resolution of ~ 7.0×1016 m−2. From the experimental result, we can observe that the line-integrated electron density in the plasma boundary region is one order smaller than that in the central region. Meanwhile, detailed variation activity of electron density can be investigated by the boundary channel. The newly developed boundary interferometry system will be helpful for the research of plasma physics of divertor discharge.

    In the near future, some other oblique channels will be developed to improve the constraint of the electron density profile reconstruction, and the interferometry system also will be upgraded into the three-wave based polarimeter/interferometer which can simultaneously detect the electron density and Faraday rotation angle.

    This study was funded and supported by the Comprehensive Research Facility for Fusion Technology Program of China (No. 2018-000052-73-01-001228), the HFIPS Director’s Fund (No. YZJJKX202301), Anhui Provincial Major Science and Technology Project (No. 2023z020004) and Task JB22001 from the Anhui Provincial Department of Economic and Information Technology.

  • [1]
    Boboc A et al 2010 Rev. Sci. Instrum. 81 10D538 doi: 10.1063/1.3478146
    [2]
    Fischer R et al 2010 Fusion Sci. Technol. 58 675 doi: 10.13182/FST10-110
    [3]
    Shibata Y et al 2011 Plasma Fusion Res. 6 1302136 doi: 10.1585/pfr.6.1302136
    [4]
    Van Zeeland M A et al 2018 Rev. Sci. Instrum. 89 10B102 doi: 10.1063/1.5037461
    [5]
    Zhuang G et al 2016 J. Instrum. 11 C02026 doi: 10.1088/1748-0221/11/02/C02026
    [6]
    Jie Y X et al 2000 Int. J. Infrared Millimeter Waves 21 1375 doi: 10.1023/A:1026496705525
    [7]
    Shi N et al 2011 Plasma Sci. Technol. 13 347 doi: 10.1088/1009-0630/13/3/14
    [8]
    Derafshi Z H, Frounchi J and Taghipour H 2010 A high speed FPGA implementation of a 1024-point complex FFT processor In: Proceedings of the Second International Conference on Computer and Network Technology Bangkok: IEEE 2010: 312
    [9]
    Huang X H, Wang Z H and Chou G Q 2007 New method of estimation of phase, amplitude, and frequency based on all phase FFT spectrum analysis In: Proceedings of the International Symposium on Intelligent Signal Processing and Communication Systems Xiamen: IEEE 2007: 284
    [10]
    Gil C et al 2008 Rev. Sci. Instrum. 79 10E710 doi: 10.1063/1.2956824
    [11]
    Ito Y et al 2005 Fusion Eng. Des. 74 847 doi: 10.1016/j.fusengdes.2005.06.241
    [12]
    Juhn J W et al 2010 Rev. Sci. Instrum. 81 10D540 doi: 10.1063/1.3492381
    [13]
    Mlynek A et al 2014 Rev. Sci. Instrum. 85 11D408 doi: 10.1063/1.4890574
    [14]
    Yao Y et al 2017 J. Instrum. 12 C12043 doi: 10.1088/1748-0221/12/12/C12043
    [15]
    Ding B G et al 2015 Plasma Sci. Technol. 17 837 doi: 10.1088/1009-0630/17/10/05
    [16]
    Lv S S et al 2021 Sensors 21 4463 doi: 10.3390/s21134463
  • Cited by

    Periodical cited type(3)

    1. Wang, S.Q., Zou, X.L., Duan, X.R. et al. Interaction between MTM and EGAM for energy and particle confinement improvements on HL-3 tokamak. Nuclear Fusion, 2025, 65(2): 026013. DOI:10.1088/1741-4326/ad9ba4
    2. Yang, Z., Zhong, W., Xia, F. et al. Implementing deep learning-based disruption prediction in a drifting data environment of new tokamak: HL-3. Nuclear Fusion, 2025, 65(2): 026030. DOI:10.1088/1741-4326/ada396
    3. Tong, R., Zhou, Y., Zhong, W. et al. A new Q-band comb-based multi-channel microwave Doppler backward scattering diagnostic developed on the HL-3 tokamak. Plasma Science and Technology, 2025, 27(1): 015102. DOI:10.1088/2058-6272/ad8c86
    1. Wang, S.Q., Zou, X.L., Duan, X.R. et al. Interaction between MTM and EGAM for energy and particle confinement improvements on HL-3 tokamak. Nuclear Fusion, 2025, 65(2): 026013. DOI:10.1088/1741-4326/ad9ba4
    2. Yang, Z., Zhong, W., Xia, F. et al. Implementing deep learning-based disruption prediction in a drifting data environment of new tokamak: HL-3. Nuclear Fusion, 2025, 65(2): 026030. DOI:10.1088/1741-4326/ada396