Online calibration of laser-induced breakdown spectroscopy for detection of heavy metals in water
-
Graphical Abstract
-
Abstract
In order to reduce the fluctuation of LIBS detection spectrum of liquid sample, the full-spectrum sum method and the internal standardization method is adopted, using an equal-RSD normalization algorithm to calibrate the detection spectrum. Experiment result shows that the full-spectrum sum method reduced the RSD of parallel samples of Cd and Cr to 9.4% and 11.06% from 28.32% and 31.93% respectively, yielded better overall calibration than the single- element internal standardization approach, thereby suggesting that the former method is convenient and effective for online calibration of LIBS for detection of aqueous heavy metals.
-
-